SCINTILLATION DETECTOR


[image]

Manufacturer: Canberra
Model: 802


Description


A Monoline Crystal Assembly which includes a high resolution NaI(TI) crystal, a photomultiplier tube, an internal magnetic/light shield and a chrome plated aluminum housing.




Specifications


Resolutionapproximately 8% at 662 keV of Cs-137
Window 0.02 in. aluminum, density 147.9 mg/cm^2
Reflectoroxide 1/16 in. thick, density 88 mg/cm^2
Magnetic/Light SheildConetic lined steel
Typical Operating Voltages
Cathode to Anode+1100 V DC
Dynode to Dynode+ 80 V DC
Cathode to Dynode+ 150 V DC