Digital Capacitance Meter


[image]

Manufacturer: Duncan Instruments
Model: DI 200C


Description


Gives a direct reading of capacitance on a 3.5 digit LCD display. Nine ranges give precision readings from 0.1 pF to 20,000 uF.




Specifications


Normal RangeMax In-range DisplayResolution
200 pF199.9 pF0.1 pF
2 nF1.999 nF1 pF
20 nF19,99 nF10 pF
200 nF199.9 nF100 pF
2 uF1.999 uF1000 pF
20 uF19.99 uF0.01 uF
200 uF199.9 uF0.1 uF
2000 uF1999 uF1 uF
20000 uF19990 uF10 uF



Click here to go to the top of this page.


Operating Instructions


  1. Observe Polarity when connecting polarized capacitors.
  2. Full discharge any capacitors.
  3. Never apply voltage to the test jacks.
  4. Set power switch to "ON" position.
  5. Select the range switch for the maximum expected capacitance.
  6. Check the zero display adjustment to read zero (limited to approximately ±20 pF).
  7. Connect the alligator clips to the capacitor leads.
  8. Read the display in the units that your range is on
  9. If the display is "1" then go to a higher range



Click here to go to the top of this page.


keywords: electrical measurement device, capacitance measurement device, capacitance meter, farad measurement device.