Manufacturer: Fluke
Model: 8000 A
Cost: $450 (1979 )


Description


Versatile meter capable of five different functions, i.e. AC and DC currents, AC and DC voltages, and resistance. Measured values are displayed on a digital readout.




Specifications


DC Voltage
Ranges ±199.9 mV to 1199 V in 5 ranges
Accuracy ±(0.1% of reading)
Input Resistance10 Mohms, all ranges
DC Current
Ranges +- 199.9 uA to 1999 mA in 5 ranges
Accuracy ± (0.3% of reading)
AC Voltage
Ranges199.9 mV to 1199 V in 5 ranges
Accuracy ±(0.5% of reading) 45 Hz to 10 kHz
±(1.0% of reading) 10 kHz to 20 kHz
Input Impedance10 Mohms in parallel with 100 pf
AC Currents
Ranges 199.9 uA to 1999 mA in 5 ranges
Accuracy ±(1.0% of reading) 45 Hz to 10 kHz, except 1999 mA range
Resistance
Ranges 199.9 ohms to 19.99 Mohms in 6 ranges
Accuracy ±(0.2% of reading) except 20 Mohm range



Click here to go to the top of this page.


Operating Instructions


  1. Turn on meter with meter disconnected from circuit.
  2. Check zero of meter as follows: connect lead across V-Ohm and Common inputs, depress DCV and 200 MV buttons, readout should read 1 digit or less.
  3. Select the appropriate range and function and then connect the meter into the circuit, ensuring the correct pair of inputs are used. If the range is not know start at highest range and work down. If display does not show all digits lit up or if display flashes a higher range should be selected.