Strip Chart Recorder


[image]

Manufacturer: Goerz Metrowatt
Model: SE120


Description


Calibrated measuring ranges and continuously adjustable intermediate ranges allow all values from 0.4 mV to 200mV, 10 mV to 250 V, 0.1 mA to 1A and temperatures from -50°C to +600°C to be recorded across the full chart width.




Specifications


Voltage Measurement
Voltage Accuracy± 0.5% of full scale
Voltage Linearity± 0.25%
Time Measurement
Accuracy± 30 ppm at 25° C
Temperature Influence± 70 ppm in the range 0 - 50° C
Input Impedance
10mv-100mV-1V ranges1 G Ohms
10V-100V ranges1 M Ohms
Miscellaneous
Input Current10 nA maximum
Maximum permissible overload250 V
Recording width250 mm
Scale0..100 linear, 20 divisions
Dead Band0.3% max.
Response Timeapprox. 0.5 s
Slewing Speedapprox. 50 cm/s
Frequency Responsea sine function of 1 Hz
DampingAccording to DIN 43 782. Overshoot or rounding response within ±1% of recording width.
Recording PaperChart rolls of 270 mm width and 32 m length with free paper run-out.



Click here to go to the top of this page.


Operating Instructions


  1. Connect the voltage to be measured to the sockets on the top of the instrument. If desired, a ground connected to the case of the instrument is on the back.
  2. The pen may be raised and lowered with the slide labelled PEN just to the right of the paper.
  3. Choose the appropriate voltage scale with the rotary switch to the right of the case. If in doubt, choose the largest possible value first.
  4. Adjust the zero voltage point with the small knob just below the voltage scale switch.
  5. The voltage scale is calibrated when the line on the small switch above the voltage scale one points to CAL. If it points to VAR then the voltage scale is continuously adjustable with the small knob in the upper right.
  6. Select the desired chart speed with the rotary knob on the left.
  7. The initial position of the chart may be adjusted with the switch labelled POS.