Click here to go to UPSCALE home page.
Click here to go to I Year Lab home page.

You can jump directly to a section of this specification sheet by selecting an item below and clicking the left button.


Universal Counter


[image]

Manufacturer: Hewlett Packard
Model: 5314 A


Description


100 MHz/100 ns Universal Counter. It features a seven-digit, seven-segment LED display with overflow indication, seven function performance, and full input signal conditioning. Seven measurement functions are: Frequency, Single-Shot Period, Period Average, Time Interval, Totalize, Ratio, and Self Check. This is accomplished by a single LSI integrated circuit. The input signal is AC coupled and can be conditioned as follows: slope selection, trigger level, and attenuation.




Specifications


INPUT Characteristics
RangeChannel A (10 Hz - 100 MHz)
Channel B (10 Hz - 2.5 MHz)
Sensitivity25 mV rms to 100/2.5 MHz. 75 mV p-p min. pulse with 5/50 ns (Channel A/Channel B).
Impedance1 M Ohms NOMINAL shunted by < 30 pF.
AttenuatorX1 or X20 NOMINAL (Channel A only)
FREQUENCY (A)
Range10 Hz to 10 MHz direct count
1 MHz to 100 MHz prescaled by 10
Resolution ± LSD
Accuracy± LSD ± (time base error) X FREQ
PERIOD (A)
Range10 Hz to 2.5 MHz
LSD displayed(100 ns) / N for N=1 to 1000
Resolution± LSD ± 1.4 X (Trigger Error) / N
AccuracyResolution ± (time base error) X PER
TIME INTERVAL (A to B)
Range250 ns to 1 s
LSD displayed100 ns
Resolution± LSD ± START Trigger Error ± STOP Trigger Error
AccuracyResolution ± (time base error) X T1.

LSD: Least Significant Digit.



Click here to go to the top of this page.


Operating Instructions


  1. Turn unit ON and set COM A/SEP switch to SEP position
  2. Connect input signal to INPUT A jack



Click here to go to the top of this page.


keywords: electrical measurement device, frequency measurement device, time measurement device.


Document created July 22, 1996. Original author: Jaime Espinoza.
Last updated 07/26/96. Revised by David Harrison.