MULTIMETER (digital)


[image]

Manufacturer: Keithley
Model: 179A
Cost: $705 (Nov. 1986)


Description


Versatile high precision meter capable of five different functions, i.e. true rms AC and DC currents, true rms AC and DC voltages, and resistance. Measured values are displayed on a digital readout.




Specifications


DC Voltage
Ranges200 mV to 1200 V in 5 ranges
Accuracy ±(0.04% of reading) on 200 mV scale
±(0.04% of reading) on other scales
Input Resistance 10 Mohms, all ranges
DC Current
Ranges 200 uA to 20 A in 6 ranges
Accuracy 2 uA through 2000 mA ± (0.02% of reading)
20 A ± (0.5% of reading)
Resistance 200 uA scale: 1 kOhm
2 mA scale: 100 Ohm
20 mA scale: 10 Ohm
200 mA scale: 1 Ohm
2000 mA scale: 100 mOhm
20 A scale: 10 mOhm
Resistance
Ranges 20 kohm to 20 Mohm in 5 ranges
Accuracy (HI) ± (0.1%) 20 MOhm scale
± (0.04%) other scales
Accuracy (LO) ± (0.15%)
Applied Voltage (HI)2V
Applied Voltage (LO)0.2V
AC Volts
Ranges 200 mV through 1000 V in 5 ranges
Accuracy ± (0.7% of reading) on 300 mV scale
± (0.6% of reading) on 2V scale
± (0.5% or reading) on other scales
Freq. range 100 Hz - 10 Hz
Input Impedance 1 Mohms shunted by less than 75 pF.
AC Current
Ranges 200 uA through 20 A in 6 scales
Accuracy ± (1% = 15 digits)
Freq. range 45 Hz - 10 kHz
Resistance 200 uA scale: 1 kOhm
2 mA scale: 100 Ohm
20 mA scale: 10 Ohm
200 mA scale: 1 Ohm
2000 mA scale: 100 mOhm
20 A scale 10 mOhm



Click here to go to the top of this page.


Operating Instructions


Obvious, except for the "Hi/Lo" switch. When used as an ohmmeter, when the switch is in the "HI" position it applies 2V across the resistance to be measured, when in the "LO" position it applies 0.2V across the resistance. The "HI" position is automatic for the 2 kOhm ranges, the "LO" position is automatic for the 20 MOhm range, and at intermediate ranges is user-selectable. The "HI" position provides greater accuracy, but the voltage is sufficient to cause unwanted effects on some components, such as forward conduction of semiconductor junctions.



Click here to go to the top of this page.


keywords: voltmeter, voltage measurement device, resistance measurement device, ohmeter, electrical measurement device, dmm, current measurement device, ammeter.


Document created October 1988. Original author: Unknown.
Last updated 01/28/05. Revised by Phil Scolieri.