SWR Meter


[image]

Manufacturer: Philips
Model: PM 7832


Description


Low-noise 1000 Hz tuned amplifier/voltmeter. Scales are calibrated in SWR and dB for use with square law detectors.




Specifications


Input frequency1000 Hz, adjustable approx. ±5% by front panel knob.
Band-width20 Hz or 100 Hz.
Sensitivity1 µV rms for full scale deflection at max band-width
Noiseat least 6 dB below full-scale with input terminated in 5000 Ohms
Inputunbiased, with a (high impedance crystals) source impedance 2k-10 k Ohms
Range70 dB in 6 steps of 10 dB and 10 dB variable by gain control
Accuracy±0.1 dB/10 dB step, max cumulative error between any two 10 dB steps ±0.2 dB. ±0.05 dB when switching NORMAL-EXPAND



Click here to go to the top of this page.


Operating Instructions


  1. Use the gain control to adjust FINE and COURSE gains.
  2. Connect instrument to the IN connector.
  3. Press the '40 dB' button.
  4. Put GAIN control in its centre position
  5. Set BANDWIDTH switch to 100 Hz.
  6. Move probe along slotted line to obtain maximum deflection
  7. Adjust GAIN so that the pointer arrives exactly at 1.0 on the VSWR scale.
  8. Move proble along slotted line to obtain minimum deflection.
  9. Read the SWR directly from the upper scale. Repeat from step 4 for scales 2 and 3.